Description The Triton Type 319 Multi-purpose CST
can be used with either 5 single radius test heads or with one multi-radii test
head for the rapid assessment of resistance to filtration.
Single
radius test heads In this mode of operation it is possible to test a
number of dosing quantities at the same time. So reducing the overall test time
or allowing for more discrete tests to be performed. In each case the Multi-CST
operates in a very similar method to the Type 304B or Type 304M CST units.
Each Type 319
CST is supplied with 1 single radius test head; any others are purchased as
separate items.
Multi-radii test head The multi-radii
test head has 5 probes set at intervals to give the same volume between each
probe. This is illustrated in the chart below. It has been shown that for a
wide range of different types of sludge there is an approximately linear
relationship between the specific resistance to filtration, using the standard
Buchner funnel apparatus and the multi-radii test head.
When using
the multi-radii test head the Type 319 unit automatically divides the time by
the probe number. Since filtration fulfils the classical parabolic relationship
between time and volume of filtrate, then the arithmetic difference in the
reading is constant and directly proportional to the specific
resistance.
The
Multi-CST is supplied with 2 boxes of filter paper as standard.
Press here to see CST video>>>
|